BAILOUT IMPLANTATION OF A NEW SINGLE-BRANCHED STENT GRAFT FOR THE AORTIC ARCH

BAILOUT IMPLANTATION OF A NEW SINGLE-BRANCHED STENT GRAFT FOR THE AORTIC ARCH

(A) Computed tomography angiogram shows aortic arch pseudoaneurysm at the level of the distal anastomosis.

(B) Three-dimensional reconstruction shows compression on the pulmonary artery

BAILOUT IMPLANTATION OF A NEW SINGLE-BRANCHED STENT GRAFT FOR THE AORTIC ARCH

Figure 2. Six-month computed tomography angiogram control. (A) Lateral view shows good positioning of the aortic arch endograft with complete exclusion of the pseudoaneurysm.

(B) Three-dimensional reconstruction shows complete aortic arch exclusion and side branch in innominate artery

Artboard 1.png

PATIENT INFORMATION

74-year-old male, COPD, previous MI, right nephrectomy (Ca), Diabetes Mellitus, Dyslipidemia, Button Bentall + Hemi Arch for Acute Type A Dissection

Artboard 5.png

REASON TO TREAT

Pseudoaneurysm of the left posterolateral side of the ascending aorta at the level of the distal anastomosis illustrating significant increase and compression of the pulmonary artery (Fig. 1). Due to risk of fistulization into the pulmonary artery, the patient was scheduled for urgent pseudoaneurysm exclusion with NEXUS™.

Artboard 4.png

DEBRANCHING

A right common carotid artery, left common carotid artery, left subclavian artery bypass was surgically performed and six days later, to allow complete recovery from the bypass surgery, the endovascular procedure was carried out.

Artboard 3.png

NEXUS IMPLANTATION

NEXUS™ was successfully deployed during an uneventful procedure with an optimal outcome lasting 150 minutes with 42 minutes of fluoroscopy.

Artboard 2.png

PROCEDURAL OUTCOME

Due to an uneventful and minimally invasive procedure, the patient was able to be discharged 7 days post op. Six month follow up CT angiogram (Fig 2) shows stable position of NEXUS™ and total exclusion of the anastomotic pseudoaneurysm.

Artboard 6.png

PHYSICIAN'S PUBLISHED COMMENTS

Cibin et al1  stated “Owing to their ready availability, singlebranch off-the-shelf aortic arch endovascular prostheses are a valid option in case of urgent and emergent cases.”

“A combined operation, bypass, and endovascular procedure at the same time, although technically feasible, would be long and with less neurologic control than a staged approach.”

1. Giorgia Cibin, MD, Augusto D’Onofrio, MD, Michele Antonello, MD, Cosimo Guglielmi, MD, Franco Grego, MD, and Gino Gerosa, MD. Bailout Implantation of a New Single-Branch Stent Graft for the Aortic Arch - Ann Thorac Surg 2020;110:e371-3

BlueGradientOnly.png

University of Padova

Division of Cardiac Surgery and Vascular Surgery

Padova, Italy

Ellipse 4.png

Professor Michele Antonello

Ellipse 4.png

Professor Augusto D’Onofrio

Ellipse 4.png

Ellipse 4.png

Consultant Vascular Surgeon

Consultant Cardiac Surgeon

EndospanLogoWhite.png

CAUTION: Investigational Device – Limited by United States law to investigational use. Endospan devices bear the CE marking of conformity.

MKJ-0005413 Rev1.0

bottom of page